Laboratory
Autoclave SN200/300/500, SQ500/810 |
Home > Products > SN200/300/500,
SQ500/810 |
| Highest
temperature of 135°C. Upward opening lid with Interlock |

Specifications
| Model*1 |
SN200/SN210 |
SN300/SN310 |
SN500/SN510 |
SQ500/SQ510 |
SQ810 |
| System |
Automatic high-pressure
steam sterilization system for laboratory use *2 |
Perfor-
mance |
Range of working temperatures |
45-135°C: 45-60°C
for preheating, 40-60°C for keep warm, 65-100°C for melting,
105-135°C for sterilization |
| Max. working pressure |
0.255MPa |
| Ambient temperature |
5 - 35°C |
| Lid |
Manual upward opening
with an interlock for safety |
| Main unit |
Heater |
100V, 1.2kW
(600W 2 units) |
100V, 1.6kW
(800W 2 units) |
100V, 1.9kW
(950W 2 units) |
100V, 2kW
(1000W 2 units) |
100V, 2kW
(1000W 2 units) |
| Exhaust valve |
An exhaust valve and
a slow release valve |
| Connection ports for optional accessories |
1/4" female thread
for, sample temp. sensor, chamber temp. sensor, and pressure sensor
(branching from the solenoid valve tubing) Total: 3 ports |
| Cooling fan |
Axial fan motor |
| Controller |
Temperature setting, control and
display |
PID control by microprocessor,
digital setting (by UP/DOWN keys) and display |
| Timer and its resolution |
0 or 1 minute to 99
hours 59 minutes in increments/decrements of 1 minute |
| Operation mode |
Instrument sterilization
course, liquid sterilization course, sterilization and keep/warm
course, melting and keep/warm course, customer-programmed course |
| Other features |
Key locking, presetting,
saving, preheating, forced cooling, sample temperature sensor (option),
memory locking, saving the history of up to 20 errors, indication
of accumulated working hours, indication of present time, on-off
beeping setting |
| Safety measures |
Sensor malfunctioning,
SSR short-circuit, broken heater wire, prevention of idle heating
(liquid expansion type), alarm against the absence of a drain bottle,
failure in locking the cover, memory malfunctioning, pressure relief
valve (0.255 MPa) |
| Pressure vessel standard |
Small-sized pressure
vessel (notification of installation is not required) |
| Standard |
External dimensions(mm) *3 |
400(W)× 530(D)× 848(H) |
460(W)× 590(D)× 848(H) |
460(W)× 590(D)× 1,058(H) |
520(W)× 660(D)× 876(H) |
520(W)× 660(D)× 1161(H) |
| Effective dimensions of chamber
(mm) |
240(inner diameter)
×445(depth) |
300(inner diameter)
×445(depth) |
300(inner diameter)
×665(depth) |
370(inner diameter)
×410(depth) |
370(inner diameter)
×750(depth) |
| Effective Capacity of chamber |
20L |
32L |
47L |
50L |
80L |
| Weight |
Approx. 65 kg |
Approx. 75 kg |
Approx. 85 kg |
Approx. 95 kg |
Approx. 105 kg |
| Power supply |
SN200:
100V-120V AC
12.5A-15A |
SN300:
100V-120V AC
16.5A-20A |
SN500:
100V-120V AC
19.5A-23.5A |
SQ500:
100V-120V AC
20.5A-24.5A |
SQ810:
200V-240V AC
10.5A-12.5A |
SN210:
200V-240V AC
6.5A-7.5A |
SN310:
200V-240V AC
8.5A-10A |
SN510:
200V-240V AC
10A-12A |
SQ510:
200V-240V AC
10.5A-12.5A |
| Power cord |
3m outside the equipment |
| Accessories |
Rack x 2
OSM-60
(dia205
xdepth204mm) |
Rack x 2
OSM-70
(dia262
xdepth204mm) |
Rack x 3
OSM-70
(dia262
xdepth204mm) |
Rack x 2
OSQ-90
(dia332
xdepth195.5mm) |
Rack x 3
MS-90
(dia332
xdepth195.5mm) |
| Vapor cup, Drain bottle,
Drainboard, Filter, Instruction manual |
Power
plug is not provided.
*1 Abbreviation (J) is Operation panel board in Japanese and (E)
is in English.
*2 The autoclaves are not for medical use.
*3 External dimensions exclude protrusions. |
Optional
accessories |
 |
In
addition to absorbent gauzes, laboratory garments and culturing instruments,
a wide selection of basket and containers for laboratory waste are
available. |
 |
Stainless
bucket |
| Product code |
Description |
Model |
Corresponding models |
|
| 241087 |
Basket (mesh type) |
OSM-60 |
For SM/SN/SP200 |
|
| 241088 |
OSM-70 |
For SM/SN300/500/SP300 |
*1 |
| 241089 |
OSM-80 |
For SM/SN/SP500 |
*1 |
| 241099 |
OSQ-90 |
For SQ500 |
|
| 241093 |
Basket (mesh type) with stacking
fitting |
OSQ-30 |
For SM/SN/SP200, with two fittings/set |
*1 |
| 241092 |
OSQ-40 |
For SM/SN/SP300, with two fittings/set |
*1 |
| 241091 |
OSQ-50 |
For SM/SN/SP500, with three fittings/set |
*1 |
| 241090 |
OSR-40 |
For SQ500, with two fittings/set |
|
| 241096 |
Basket (mesh type) with a mesh shelf |
OSQ-60 |
For SM/SN/SP200, with 1 mesh shelf |
*1 |
| 241095 |
OSQ-70 |
For SM/SN/SP300, with 1 mesh shelf |
*1 |
| 241094 |
OSQ-80 |
For SM/SN/SP500, with 2 mesh shelves |
*1 |
| 241097 |
OSR-50 |
For SQ500, with 1 mesh shelf |
|
| 241083 |
Stainless solid basket |
OSR-10 |
For SM/SN/SP200 |
*1 |
| 241084 |
OSR-20 |
For SM/SN/SP300/500 |
*1 |
| 241098 |
OSR-60 |
For SQ500 |
|
| 241150 |
Stainless bucket |
OSN10 |
For SM/SN/SP200 |
|
| 241151 |
OSN12 |
For SM/SN/SP300/500 |
|
| 241152 |
OSN14 |
For SQ500 |
|
| 241153 |
Chamber temperature sensor |
OSR-70 |
For SN/SQ, Type T thermocouple, 3 pieces/set |
|
| 241154 |
Sample temperature sensor |
OSN16 |
For SN/SQ, Type T thermocouple, 1 piece |
|
| 241155 |
Pressure gauge |
OSN18 |
For SN/SQ, external installation |
|
| 241156 |
Recorder |
RM10C |
6-point type, temp. and pressure recording external
installation |
|
| 241157 |
External alarm output terminal |
OSN20 |
For SN/SQ |
|
| 241158 |
Temperature output terminal |
OSN22 |
For SN/SQ, 4 - 20 mA |
|
| 241159 |
Time-up output terminal |
OSN24 |
For SN/SQ, relay, contact output |
|
| *1: Can be also used with conventional
models SM200/300/500/510 and SE200/300/500/510. |
|