| Model |
GAS410 |
| Solvent
recovery system |
Capacitor + freezer |
| Performance |
Circulating gas |
N2 gas (sealed circulation
when connected to ADL311S-A or GB210-A) |
| Circulating wind amount |
0.12 to 0.65m3/min |
| Configuration |
Compressor (for spraying) |
Linear compressor integrated |
| Circulation blower |
Roots blower |
| Solvent recovery container |
2L flask |
| Freezer |
Air-cooled condensation
full-sealed type: 400W R404A |
| Solvent recovery mechanism |
Capacitor cooling mechanism |
| Filter |
Cartridge filter |
| Instruments |
Cooling trap temperature
display monitor |
| Filter differential
pressure meter (monitor for clogging of filter) |
| O2 density display
monitor |
| Blower wind amount
adjusting volume |
| O2 sensor |
Solid electrolyte (Zirconium)
limit current type |
| Pump |
For circulation for
O2 sensor |
| Safety
unit |
O2 density meter, flammable
gas alarm, electric leakage breaker, N2 gas forced introduction (when
removing nozzles) |
| Standards |
External size |
W700 x D950 x H1,500
mm |
| Weight |
Approx. 130 kg |
| Power supply (50/60
Hz) rated current |
AC200 to 240V 5A (15A) |
| Required N2 amount |
15 L/h at 0.1 MPa |
| Accessories |
Set of connection parts,
anti-seismic clamps, interface cable, sample gas for gas alarm inspection,
2L flask |