DE410 |
DE610 |
DT410 |
DT610 |
DE410U |
DT610U |
| Performance |
| Circulation method |
Forced air circulation |
| Operating temperature range |
Room temperature +30°C to 260°C |
Room temperature +30°C to 360°C |
Room temperature +30°C to 200°C |
| Temp. adjustment accuracy |
±0.3°C(at 260°C) |
±0.3°C(at 360°C) |
±0.3°C(at 200°C) |
| Temp. uniformity |
±2.5°C(at 260°C) |
±4.0°C(at 360°C) |
±4.0°C(at 200°C) |
| Max. temp. reaching time |
Approx. 70 min.
(to 260°C) |
Approx. 60 min.
(to 260°C) |
Approx. 80 min. (to 360°C) |
Approx. 60 min. (to 200°C) |
| Clean level |
Class 100 (when the temperature is stable) |
Class 100 (when the temperature is
stable, as well as is going up or down) |
| Components |
| Interior Material |
Stainless steel SUS 304 |
| Exterior |
Cold rolled steel plate with melamine resin baking finish |
| Heat insulating material |
Glass wool |
Ceramic fiber |
Glass wool |
| Stainless pipe heater |
2.5kW |
3.6kW |
3.6kW |
5.2kW |
2.5kW |
3.6kW |
| Fan type |
Scirocco fan, Condenser type motor 400W |
| Differential pressure meter |
Analog type (0 to 500 Pa) |
| Cable hole |
30 mm I.D. (right side) |
| Additional functions |
Exhaust damper (manual type) 61mm O.D. |
| Filter |
Heat resistant HEPA filter (dust collect : more than 99.97% at 0.3μm ) |
High performance HEPA filter (dust
collect : more than 99.97% at 0.3μm) |
| Controller |
| Temp. controller |
PID control by micro processor |
| Temp. setting method |
Digital setting with  key |
| Temp display method |
Digital display |
| Other display |
Operation monitor (operation condition graphic display by LED patterns) |
| Timer |
1 min. to 99 Hrs. 59 min. and 100 Hrs. to 999 Hrs. |
| Operation function |
Fixed temp., Auto-start, Auto-stop, Program operation (16 segments) |
| Additional functions |
Timer (max. 24Hr.), Integration (max. 49999Hr.), Calibration off-set, time |
| Heater circuit control |
Triac zero-cross control |
| Temp. sensor |
K-thermocouple (double sensor) |
| Safe Device |
| Safety countermeasures |
Self diagnosis functions (sensor, heater, triac, automatic overheating prevention), Key lock, Independent
overheating prevention, Door switch, Electric leakage breaker |
| Standard |
Internal dimensions (W×D×Hmm)
External dimensions(W×D×Hmm)
Internal capacity |
450×450×450
700×1,003×1,765
91 L |
600×600×600
850×1,153×1,765
216 L |
450×450×450
700×1,003×1,765
91 L |
600×600×600
850×1,153×1,765
216 L |
450×450×450
700×1,003×1,765
91 L |
600×600×600
850×1,153×1,765
216 L |
| Shelf plate with standard load |
Approx. 30 kg / piece |
| Shelf rest step number / pitch |
12 steps /30mm |
17 steps /30mm |
12 steps /30mm |
17 steps /30mm |
12 steps /30mm |
17 steps /30mm |
| Power source 50/60Hz |
200V
three phase,
9.5A |
200V
three phase,
12.5A |
200V
three phase,
12.5A |
200V
three phase,
17.5A |
200V
three phase,
9.5A |
200V
three phase,
12.5A |
| Weight |
Approx. 220 kg |
Approx. 270 kg |
Approx. 220 kg |
Approx. 270 kg |
Approx. 220 kg |
Approx. 270 kg |
| Accessories |
| Shelf plate |
Stainless steel, 2 pcs |
Stainless steel, 3 pcs |
Stainless steel, 2 pcs |
Stainless steel, 3 pcs |
Stainless steel, 2 pcs |
Stainless steel, 3 pcs |
| Shelf bracket |
4 pcs |
6 pcs |
4 pcs |
6 pcs |
4 pcs |
6 pcs |
| Optional Accessories |
| Additional shelf plate / Analog recorder / Auto damper / Emergency stop switch / Exhaust duct etc. |