Yamato Scientific America, Inc.
Search
Site Map | Contact Us
SINCE 1889 What's New About Yamato Products Technical Warranty & Repair HOME
Pulvis Mini Spray / Mini Bed series GA22    Home > Products > GA22

Processing Capacity
50g to 300g
Operating Temperature Range
40°C to 200°C
Liquid Sample Flux
up to 28 ml/min.
Spraying Nozzle
for Liquid and Air

A system to granulate powder of small amount and to dry moist powder.

The GA22 uses a fluid bed and it can be used to granulate powders or to dry moist powders.
It is capable of processing samples of between 50g and 300g and well suited for expensive samples and laboratory experiments.

  • The setting of testing conditions is easily entered from the operation panel.
  • The testing mode can be easily observed as the chamber is made of superhard glass.
  • Since it is equipped with impellers, consistently good fluid conditions are obtained.


* Control panel
GA22 Control Panel
*
Inlet temperature, outlet temperature and drying air volume are displayed digitally. Condition setting dials and stage level control switch are located on the front panel for easy use.

GA22
* Since a small quantity of samples from 50g to 300g can be treated this device is most appropriate for the treatment of expensive samples and the experiments at the laboratory level.

GA22

GA22
GA22

* Easy operation and maintenance
GA22 GA22

* A stirring blade is attached to the perforated plate so that inferior fluid condition can be improved and wet powdered matters drying can be processed by stirring.

* The stage level control switch are located on the front panel for easy use.
Cleaning of chamber is also made very easy.

 
* Spraying nozzle
Nozzle size (µm)
Model 1A 1 2A 2 3
Nozzle No. 1650 2050 2050 2850 2850
(F) A 406 508 508 711 711
B 270 1270 1270 1270 1270
Nozzle No. 64 64 70 70 64
(A) C 1626 1626 1778 1778 1626

Model 1A : Standard nozzle
GA22 Liquid nozzle (F)
GA22 Liquid nozzle
Air nozzle (A)
GA22 Air nozzle
 
* System diagram
GA22 System diagram



* Specifications

 

GB22 (Basic Unit)

Structure

Heater

2 kW

Aspirator

Bypass-type blower

Sample liquid feed pump

Quantitative peristaltic pump, flow rate variable up to 28ml/min.

Stirring system

Using an induction motor

Compressed air blowing system

Pulse jet system, an electromagnetic valve and electronic timer

Temperature control range

40°C to 200°C

Temperature control accuracy

±1°C

Temperature setting and display method

Digital

Dry air flow meter

Digital display

Spray pressure gauge

Bourdon tube pressure gauge, measuring range : 0 to 0.3MPa (0 to 3kg/cm2)

Standard

Power source (50/60Hz)

AC 200/220V, single phase
14A/12.7A

External dimensions(W×D×Hmm)

760×420×1,345

Weight

Approx. 110kg

 

 

GF22 (Pulvis Mini Bed Attachment)

Structure and Standard

Processing capacity

50 to 300g (batch method)

Fluid bed chamber capacity

3L

Spray nozzle

Two-liquid nozzle used

Stirring blade

Incorporated in the fluid bed chamber (using a stirring system of Model GA22)

Filter

Polyester (carbon fiber mixed PTFE membrane laminate)

Filter cleaning

Pulse jet system (using a compressed air blowing system of Model GB22)

Glass parts

Super-hard glass

Weight

Approx. 13kg

CONTACT US Toll free 1-800-2-YAMATO/ E-mail:customerservice@yamato-usa.com