Model |
DN410H |
DN610H |
Performance |
Convection Type |
Forced Convection |
Operating Temp. Range |
Room Temp. +10 to 360°C |
Temp. Adjustment Accuracy |
±0.2°C (at 360°C) |
Temp. Distribution Accuracy |
±3.0°C (at 360°C) |
Time to Reach to Max. Temp. |
Approx. 60min. |
Components |
Interior / Heat Insulation Material |
Stainless Steel SUS304 / Glass Wool + Ceramic Fibre |
Heater |
SUS Pipe Heater 3.0kW |
SUS Pipe Heater 4.0kW |
Fan |
Sirocco Fan, Condenser Type Motor 30W |
Cable Hole/Exhaust Hole |
I.D. 30mm (Right Side) / I.D. 30mm x 2 (Back Side) |
Temp. Control Type |
PID Control by Micro Computer |
Temp. Setting Type |
Digital Setting by  Key |
Temp. Display |
Digital Indication by Green LED |
Other Display |
Temp. Pattern for Operating Display by LED |
Timer / Timer Resolution |
1min. to 99hrs. 59min. or 100hrs. to 999hrs. / 1min. or 1hrs. |
Operating Function |
Fixed Temp. Operation, Automatic Start, Automatic Stop, Program Operation (Max. 16 Segments, Splitable, Repeating, Gradient Operation) |
Additional Function |
Timer Function (Actual Timer within 24hrs.), Total Operation Hours Timer (to49,999hrs.), Calibration Off Set Function, Clock Display (Current Time Display, Displayable at the Power Off) |
Heater Circuit Control / Sensor |
Triac Zero-cross Type / K-Thermocouple (Double Sensor) |
Safety Device |
Safety Countermeasure |
Self Diagnosis Function (Sensor, Triac, Main Relay, Heater, Automatic Overheat Prevention), Key Lock Function, Door Switch, Independent Overheat Prevention, Electric Leakage Breaker |
Standard |
Internal Dimensions |
W470 x D450 x H450mm |
W620 x D600 x H600mm |
External Dimensions |
W640 x D695 x H915mm |
W790 x D846 x H1,065mm |
Internal Capacity |
95L |
223L |
Shelf Max. Load |
Approx. 30kg/pc. |
Shelf Support Qty./Pitch |
12pcs. / 30mm |
17pcs. / 30mm |
Power |
1-Phase, 200V, 50/60Hz, 16A |
1-Phase, 200V, 50/60Hz, 21A |
Weight |
Approx. 80kg |